Method | How the input signal is measured or generated:- Controlled: The calibrator controls an external device, like a pressure controller, but the generated signal is measured with the calibrator
- Controlled and measured: The calibrator controls an external device and gets the measurement readings from the external device.
- Keyed: The measurement result is keyed, i.e. manually entered into the calibrator
- Measured: The calibrator measures the instrument’s input signal
- Sourced: The calibrator generates or or simulates the instrument’s input signal
|
Port | Which of the calibrator's RTD/TC ports is going to be used in the calibration. Valid for measurements and simulations done using RTD/TC ports. Note, that different calibrators have different amount of RTD/TC channels. (e.g. MC2 has just one RTD channel and MC6-T has 3). |
RJ Mode | Selection is valid for thermocouples, when input methods is set to Controlled, measured or Sourced. Select a suitable reference junction compensation method for your calibration: - External R1 or R2: Use External RTD sensor to measure the reference junction temperature. When using MC2 or MC4, only External R1 can be used.
- Internal: Use calibrator's automatic reference junction compensation capability.
- Manual: Reference junction is in a fixed, known temperature.
- Fixed (0°C): The reference junction is in an ice bath.
Read more about reference junction compensation from Beamex blog |
External sensor type | Select the RTD sensor type used to measure the reference junction temperature. Field is shown only when RJ mode is set to External R1 or R2. |
RJ value | Set the known temperature of the reference junction in °C. Available only when RJ Mode has been set to manual |
Sensor supply | Valid for pressure. Select this if the pressure sensor requires supplied by the calibrator. Available selections are: - Undefined: Sensor supply not in use
- Current
- Voltage
|
Sensor supply level | This field is visible only if sensor supply is set to Current or Voltage . You can set the value for the sensor supply here. The value is validated based on the selected calibrator model and supply quantity. |
Frequency amplitude (V) | Field is available only when the input quantity is Frequency and method is sourced. Amplitude of the frequency generation is set here. |
Wave form | Field is available only when the input quantity is Frequency and method is sourced. Available values are: |
Contact type | Available for frequency functions. Available values are: - Dry contact (low supply):
- Wet contact:
|
Frequency pulse duty cycle (%) | Set the pulse duty cycle, between 1 ... 99 %. Field valid only with input quantity is Frequency and method is set to sourced. |
Trigger level | Field is shown only when input = frequency, method = measured and contact type wet contact |